FEATURES:
The performance and functional characteristics of YH-600B Pro include all the user-friendly features of the predicate device. YH-600B Pro is powered by one 1.5 V AA battery. Testing was performed to demonstrate that the performance of YH-600B Pro in its intended environment is as safe and effective as that of the legally marketed predicate devices. The safety and effectiveness of YH-600B Pro have been verified through performance-related testing that consisted of Electrical Safety, Electromagnetic Compatibility, Mechanical and Environment Testing. The recorded data, such as SpO2, pulse rate, airflow waveform, snore and CPAP pressure, can be replayed on a computer via the PolyLogic Analysis Software which is provided with YH-600B Pro. The device can be used as a standalone unit for recording data. It can also interact with a commercially available IEC60950-1 compliant computer, which can be used to review data. The device does not have any audible alarms, and, therefore, should not be used for continuous monitoring. Patient is an intended operator, who can safely use all functions and can perform all maintenances mentioned in the User Manual. All parts of the device are applied parts.
Monitoring Parameters:
SpO2, Pulse Rate, Airflow Waveform, Snore, CPAP Pressure, Wrist Movement (Sleep / Awake), Thoracic / Abdominal Respiratory Effort, Body Position
Exclusive software, simple to operate and analyse!
Technical Specifications
General Specifications
SpO2 Range
|
0%~100%
|
Pulse rate Range
|
30~250 bpm
|
Power supply
|
DC 1.5V ± 0.15V (1×1.5V R6 AA)
|
Minimum record duration
|
No less than 8 hours
|
Data storage
|
64 records can be stored on TF card
|
Dimensions
|
78 mm × 60 mm × 21 mm (main device)
|
Weight
|
75g (main device without battery)
|
Type of protection against electrical shocks
|
Internal power supply
|
Classification of protection against electrical shocks
|
Type BF applied part
|
Waterproof rating
|
IP22
|
Working mode
|
Continuous operation
|
Wavelength
Red
|
663 nanometers
|
Infrared
|
890 nanometers
|
Maximal Optical Output Power
|
Less than 1.5 mw maximum average
|
Airflow waveform Frequency
|
10~40 times per minute
|
CPAP pressure Rang
|
4~20 hPa
|
Snore
|
150~300 Hz
|
Temperatures
Operation
|
5 °C~35 °C
|
Storage/Transportation
|
-20 °C~55 °C
|
Wrist movement (sleep/wake)
|
Automatic sleep-awake identification from wrist movement
|
Humidity
Operation
|
< 80% (non-condensing)
|
Storage/Transportation
|
< 93% (non-condensing)
|
Atmospheric Pressure
Operation
|
860~1060 hPa
|
Storage/Transportation
|
700~1060 hPa
|